TOP > More Detailed Information of Products > Medical Devices (Asahi Intecc Brand Products) > PTA > ASAHI Corsair Armet
Super SHINKA-Shaft
- Supportive catheter body increases the push force of the guide wire
- Low crossing profile – 0.75 mm (2.3 Fr)
- Compression resistant catheter body maintains inner lumen integrity in highly calcified lesions
- Bi-directional rotation to cross more resistant lesions
- Rotation force – 37% more torque force than Corsair
- Low entry profile 0.5 mm (0.020”)
- Metal tip resistant to trapping
The microcatheters were advanced into the calcified lesion model.
The metal tip resists deformation when Corsair Armet enters a hard lesion.
Comparative assessment when Corsair Armet measured value is “1”.
Iliac artery model was used. Microcatheters were advanced in a contra-lateral approach.
Mechanical push force was applied to proximal shaft of microcatheters and output force was measured.
All the data were obtained by company standardized test, which may differ from industry standardized tests.
All the data do not guarantee that all devices have exactly the same performance with the samples used for tests.
Recommended guide wire-0.36mm (0.014”)
Product | Catalog No. | Outer diameter | Inner diameter | Usable length | Coating length | |||
---|---|---|---|---|---|---|---|---|
Tip | Distal shaft |
Proximal shaft |
Tip | Shaft | ||||
ASAHI Corsair Armet |
WCSAR060-14N | 0.50mm (1.5Fr) |
0.75mm (2.3Fr) |
0.83mm (2.5Fr) |
0.42mm (0.017inch) |
0.48mm (0.019inch) |
60cm | 23cm |
WCSAR090-14N | 0.50mm (1.5Fr) |
0.75mm (2.3Fr) |
0.83mm (2.5Fr) |
0.42mm (0.017inch) |
0.48mm (0.019inch) |
90cm | 30cm | |
WCSAR135-14N | 0.50mm (1.5Fr) |
0.75mm (2.3Fr) |
0.83mm (2.5Fr) |
0.42mm (0.017inch) |
0.48mm (0.019inch) |
135cm | 70cm | |
WCSAR150-14N | 0.50mm (1.5Fr) |
0.75mm (2.3Fr) |
0.83mm (2.5Fr) |
0.42mm (0.017inch) |
0.48mm (0.019inch) |
150cm | 115cm |
“ASAHI”, “Caravel”, “Corsair” and “Armet” are trademarks or registered trademarks of ASAHI INTECC CO., LTD. in Japan and other countries.